PT100 / PT1000 RTD Calculator

Calculate platinum RTD resistance or temperature using the IEC 60751 Callendar–Van Dusen equation, with tolerance range.

Typical resistance
Minimum
Maximum
Input reading

About platinum RTDs

T ≥ 0 °C: R = R₀ × (1 + A·T + B·T²)

T < 0 °C: R = R₀ × (1 + A·T + B·T² + C·(T − 100)·T³)

PT100 (R₀ = 100 Ω) and PT1000 (R₀ = 1000 Ω) are platinum resistance temperature detectors. This uses the IEC 60751 Callendar–Van Dusen coefficients (with a 3920 ppm/°C option). Switch the mode to solve resistance from a temperature, or temperature from a measured resistance; the tolerance gives the resulting min/max band.